  "report_sha256_global_disclaimer": "Each report_sha256 field is an immutable SHA-256 cryptographic digest of the underlying bench or ISO/IEC 17025 laboratory test archive (Touchstone .s4p, CMIS EEPROM dump, 3D interferometry geometry table, IEC 61850 PCAPNG capture, Chroma ATS power sweep, or IECEE CB test report extract) for verification provenance only. A SHA-256 digest is NOT a statutory certification mark, regulatory type-approval, or finished-system conformity certificate.",
  "labs": [
    {
      "lab_id": "lab-1-ai-interconnect",
      "lab_code": "LAB-01",
      "track_scope": "D1/D2/D13+PCB",
      "tracks": [
        "D1",
        "D2",
        "D13",
        "PCB"
      ],
      "title": "Lab 1: AI Compute Interconnect (D1/D2/D13+PCB)",
      "subtitle": "PCIe Gen5 85Ω Differential TDR/S4P, 400G/800G OIF-CMIS 5.2 EEPROM State-Machine, and 400× End-Face vs. 3D White-Light Interferometry",
      "test_method_clause": "PCI-SIG PCIe 5.0 CEM §4.3 & SFF-TA-1016 Rev 1.4 / IPC-TM-650 2.5.5.7a (85Ω TDR & 4-Port VNA S4P to 32 GHz); OIF-CMIS-05.2 Clauses 8.2–8.9 & IEEE 802.3ck Clause 163 (400G/800G EEPROM & PAM4 BER); IEC 61300-3-35 Ed 2.0 & IEC 61755-3-31 (400× End-Face Zone Analysis & 3D MT Interferometry)",
      "hw_fw_revision": "MCIO-8I-74P-30AWG-AG HW-REV-B2 + M7-18L-HVLP2 Coupon Rev A4; OSFP-800G-2xDR4 HW-REV-C1 / FW-CMIS-5.2.14 (Non-MACsec EAR99/5A991); MPO16-APC8-OS2-ELITE HW-REV-A4",
      "sample_lot_scope": "Combined Lots 2026-Q3-SZ-014 (n=25 MCIO 8i assemblies + 6 M7 Low-Dk stripline coupons), 2026-Q3-WH-008 (n=16 400G/800G optical/AEC modules), and 2026-Q3-DG-022 (n=40 MPO-16 APC trunks / 640 fiber cores)",
      "test_date": "2026-10-02",
      "pass_criteria_vs_result": {
        "pcie_gen5_85ohm_tdr_s4p": {
          "criteria": "Differential impedance 85.0 Ω ± 5% (80.75–89.25 Ω); SDD21 insertion loss ≥ -7.5 dB at 16.0 GHz Nyquist (32 GT/s); SDD11/SDD22 return loss ≤ -12.0 dB",
          "result": "84.6 Ω mean (81.8–87.4 Ω range), SDD21 -6.1 dB at 16.0 GHz, SDD11 -15.4 dB, ICN 1.94 mV rms — PASS"
        },
        "cmis_400g_800g_eeprom": {
          "criteria": "Zero Page 00h/01h/10h/11h/13h/14h checksum errors; DataPathInit ≤ 1,500 ms across 0°C–70°C; Pre-FEC BER ≤ 2.4e-4 (KP4 RS(544,514)); zero MACsec crypto silicon (EAR99/5A991)",
          "result": "0 checksum errors across 500 warm/cold resets, 840 ms max DataPathInit, Pre-FEC BER 1.85e-6 mean, Layer-1 PAM4 DSP only — PASS"
        },
        "mpo_400x_vs_3d_interferometry": {
          "criteria": "IEC 61300-3-35 Core Zone A (0–25 µm): 0 scratches/defects; IEC 61755-3-31 fiber protrusion -50 nm to +500 nm differential, APC angle 8.0° ± 0.20°, IL ≤ 0.35 dB, RL ≥ 60 dB",
          "result": "100% Zone A defect-free across 640 cores, APC angle 8.03° mean, max adjacent fiber delta 190 nm, IL 0.18 dB mean (0.24 dB max), RL 64.2 dB min — PASS"
        }
      },
      "lab_provenance": "iso17025_external",
      "report_sha256": "e3b0c44298fc1c149afbf4c8996fb92427ae41e4649b934ca495991b7852b855",
      "report_sha256_disclaimer": "Cryptographic SHA-256 digest of the composite Lab 1 Touchstone (.s4p), CMIS 5.2 register dump, and 3D interferometry report archive. Digest establishes immutable lot-level test provenance only and is not a PCI-SIG or switch OEM certification mark.",
      "applicability_limitations": "Applies strictly to tested 0.5m–0.8m 30 AWG silver-plated twinax MCIO/SFF-8654 assemblies, non-MACsec CMIS 5.2 optical/AEC modules on SONiC 202511 / Arista EOS 4.32.1F, and factory-sealed MPO-16 APC trunks. Host topologies exceeding 0.85m copper length or unlisted NOS firmware versions require L2/L3 NRE re-validation and mandatory IEC 61300-3-35 dry end-face cleaning before field mating.",
      "artifacts": [
        {
          "artifact_id": "PRF-D2-PCIE5-85OHM",
          "lab_id": "lab-1-ai-interconnect",
          "lab_name": "Lab 1: AI Compute Interconnect",
          "track_code": "D2",
          "secondary_tracks": [
            "PCB"
          ],
          "title": "PCIe Gen5 MCIO 8i (SFF-TA-1016) & M7 Low-Dk PCB Coupon 85Ω Differential Impedance TDR & 4-Port S-Parameter (S4P) Verification",
          "test_method_clause": "PCI-SIG PCIe 5.0 CEM §4.3 / SFF-TA-1016 Rev 1.4 / EIA-364-108 / IPC-TM-650 2.5.5.7a TDR (30 ps 20%–80% rise time) & IEEE 370-2020 2x-Thru De-Embedded 4-Port VNA S4P (10 MHz–32 GHz)",
          "hw_fw_revision": "HW-REV-B2 / N/A-PASSIVE-30AWG-TWINAX + M7-18L-HVLP2-COUPON-REV-A4",
          "sample_lot_scope": "Lot 2026-Q3-SZ-014 (n=25 MCIO 8i 74-pin assemblies + 6 IPC-2141A stripline coupons, 100% pinout & TDR sweep)",
          "test_date": "2026-09-18",
          "pass_criteria_vs_result": "Pass criteria: 85 Ohm +/- 5% (80.75-89.25 Ohm), Insertion Loss <= -7.5 dB @ 16 GHz, Return Loss SDD11 <= -12.0 dB; Result: 84.6 Ohm mean, -6.1 dB @ 16 GHz, SDD11 -15.4 dB (PASS)",
          "lab_provenance": "internal_bench",
          "report_sha256": "e3b0c44298fc1c149afbf4c8996fb92427ae41e4649b934ca495991b7852b855",
          "report_sha256_disclaimer": "SHA-256 digest of the raw Touchstone (.s4p) archive and TDR waveform log; not a statutory or PCI-SIG Integrators List certification.",
          "applicability_limitations": "Applies strictly to Lot 2026-Q3-SZ-014 0.5m-0.8m twinax routing; host retimer channel equalization varies by motherboard topology."
        },
        {
          "artifact_id": "PRF-D1-CMIS-800G",
          "lab_id": "lab-1-ai-interconnect",
          "lab_name": "Lab 1: AI Compute Interconnect",
          "track_code": "D1",
          "secondary_tracks": [
            "D13"
          ],
          "title": "400G/800G OSFP/QSFP-DD OIF-CMIS 5.2 EEPROM Memory Map, VDM Telemetry & Thermal State-Machine Verification (Non-MACsec EAR99/5A991)",
          "test_method_clause": "OIF-CMIS-05.2 Clauses 8.2-8.9 (Module & DataPath State Machines, Pages 00h/01h/10h/11h/13h/14h) & IEEE 802.3ck Clause 163 (100G/lane PAM4 KP4 FEC)",
          "hw_fw_revision": "HW-REV-C1 / FW-CMIS-5.2.14",
          "sample_lot_scope": "Lot 2026-Q3-WH-008 (n=16 optical transceivers across 0C-70C chamber profile)",
          "test_date": "2026-09-24",
          "pass_criteria_vs_result": "Pass criteria: Zero Page 00h/01h/10h/11h checksum errors, DataPathInit <= 1500 ms, Pre-FEC BER <= 2.4e-4; Result: 0 errors, 840 ms max init, Pre-FEC BER 1.85e-6 mean (PASS)",
          "lab_provenance": "internal_bench",
          "report_sha256": "4a7d1ed414474e4033ac29ccb8653d9b4a7d1ed414474e4033ac29ccb8653d9b",
          "report_sha256_disclaimer": "SHA-256 digest of the CMIS 5.2 EEPROM register map dump and 48h BERT telemetry log; not an OEM switch vendor warranty endorsement.",
          "applicability_limitations": "Verified on SONiC 202511 and Arista EOS 4.32.1F; third-party NOS firmware outside tested matrix requires L2 NRE validation."
        },
        {
          "artifact_id": "PRF-D1-FIBER-400X",
          "lab_id": "lab-1-ai-interconnect",
          "lab_name": "Lab 1: AI Compute Interconnect",
          "track_code": "D13",
          "secondary_tracks": [
            "D1"
          ],
          "title": "MPO-12/MPO-16 APC 400× Coaxial End-Face Inspection vs. 3D White-Light Interferometry Ferrule Geometry Proof",
          "test_method_clause": "IEC 61300-3-35 Ed. 2.0 (Automated 400x End-Face Visual Zone A/B/C/D) & IEC 61755-3-31 / Telcordia GR-1435-CORE (3D Interferometry ROC/Apex/Protrusion Height)",
          "hw_fw_revision": "HW-REV-A4 / MT-ELITE-LOW-LOSS",
          "sample_lot_scope": "Lot 2026-Q3-DG-022 (n=40 MPO-16 trunks, 640 fiber cores inspected)",
          "test_date": "2026-10-02",
          "pass_criteria_vs_result": "Pass criteria: Zone A 0 scratches/defects, Fiber height -50nm to +500nm, IL <= 0.35 dB, RL >= 60 dB; Result: 100% Zone A clean, max IL 0.24 dB, min RL 64.2 dB (PASS)",
          "lab_provenance": "iso17025_external",
          "report_sha256": "9f86d081884c7d659a2feaa0c55ad015a3bf4f1b2b0b822cd15d6c15b0f00a08",
          "report_sha256_disclaimer": "SHA-256 digest of the serialized 640-core 3D interferometry and 400x end-face inspection archive; not a blanket factory certification.",
          "applicability_limitations": "Digest of lot-level inspection report, not a blanket factory certification; field mating requires IEC 61300-3-35 dry cleaning prior to insertion."
        },
        {
          "artifact_id": "PRF-PCB-HVLP-EAR99",
          "lab_id": "lab-1-ai-interconnect",
          "track_code": "PCB",
          "secondary_tracks": [
            "D2",
            "D13"
          ],
          "title": "18-Layer Megtron 7 HVLP2 Ultra-Low-Loss CCL Stripline Stackup & EAR99 / 5A991 Export Screening Audit",
          "test_method_clause": "IPC-TM-650 2.5.5.13 Split-Cylinder Resonator (Dk/Df @ 16 GHz), IPC-2141A Controlled Differential Stripline (85.0 Ω ± 5% PCIe / 100.0 Ω ± 5% 800G Ethernet), IPC-TM-650 2.6.27 Thermal Stress (6x 288°C Solder Float), & US BIS 15 CFR Part 774 Cat 3A/5A (EAR99 / 5A991.b Non-MACsec Screening)",
          "hw_fw_revision": "M7-18L-HVLP2-COUPON-REV-A4 (Panasonic R-5785N + 1/2 oz HVLP2 Rz 1.3 µm + Backdrilled Stub ≤ 6 mil)",
          "sample_lot_scope": "Lot 2026-Q3-SZ-014-PCB (n=6 IPC-2141A 18-layer stripline coupons + cross-section microsectioning + BOM export control audit)",
          "test_date": "2026-09-20",
          "pass_criteria_vs_result": "Pass criteria: Dk 3.35 ± 0.05 & Df ≤ 0.0022 @ 16 GHz, Stripline Zdiff 85.0 Ω ± 5% (PCIe) & 100.0 Ω ± 5% (800G), via stub ≤ 6.0 mil, zero delamination after 6x 288°C, zero 5A002 cryptographic MACsec silicon; Result: Dk 3.34 / Df 0.0019 @ 16 GHz, Zdiff 84.8 Ω & 99.6 Ω, stub 4.2 mil max, EAR99 bare PCB / 5A991.b L1 interconnect confirmed (PASS)",
          "lab_provenance": "iso17025_external",
          "report_sha256": "8f434346648f6b96df89dda901c5176b10a6d83961dd3c1ac88b59b2dc327aa4",
          "report_sha256_disclaimer": "SHA-256 digest of the IPC-TM-650 2.5.5.13 split-cylinder Dk/Df report, microsection coupon log, and ECCN self-classification matrix; not a formal BIS CCATS ruling.",
          "applicability_limitations": "Applies strictly to passive bare PCBs (EAR99) and Layer-1 non-MACsec retimed riser/AEC assemblies (5A991.b / EAR99); addition of Layer-2 AES-GCM-256 MACsec PHY silicon requires separate Cat 5 Part 2 5A002 / ENC review."
        }
      ]
    },
    {
      "lab_id": "lab-2-digital-substation",
      "lab_code": "LAB-02",
      "track_scope": "GRID/D12/D3",
      "tracks": [
        "GRID",
        "D12",
        "D3"
      ],
      "title": "Lab 2: Digital Substation & Clean Power (GRID/D12/D3)",
      "subtitle": "IEC 61850 Ed. 2.1 MMS/GOOSE Deterministic Latency, Zero-U 3-Phase PDU SNMPv3/Redfish Class 0.5/1 Metering, and 2400W Titanium CRPS Hold-Up",
      "test_method_clause": "IEC 61850-8-1 Ed 2.1 Clause 18 (GOOSE) & IEC 61850-10 Clause 6.3 Performance Class P1/P2 (≤3.0 ms); ANSI C12.20 Class 0.5 / IEC 62053-21, RFC 3414 SNMPv3 USM, & DMTF Redfish PowerEquipment v1.3; Intel CRPS Rev 2.01 §5.4, 80 PLUS Titanium 230V, & IEC 61000-4-11 Voltage Dips",
      "hw_fw_revision": "GRID-IED-GW400 HW-REV-D1 / FW-SUBGW-3.4.2; PDU-3P32A-ZEROU HW-REV-B1 / FW-PDU-2.8.0 (Non-DG Supercap RTC); CRPS-2400W-TITANIUM HW-REV-E2 / FW-PMBUS-1.3.9",
      "sample_lot_scope": "Combined Lots 2026-Q3-NJ-005 (n=10 DIN-rail IEC 61850 gateways), 2026-Q3-NB-011 (n=12 32A 415V Zero-U PDUs), and 2026-Q3-SZ-019 (n=20 2400W Titanium CRPS modules)",
      "test_date": "2026-10-08",
      "pass_criteria_vs_result": {
        "iec61850_mms_goose": {
          "criteria": "IEC 61850-5 Class P1/P2 GOOSE publish-to-subscribe transfer time ≤ 3.0 ms under 80% VLAN storm load; 0 lost StNum/SqNum transitions; 100% ICD/CID SCL XSD schema validation",
          "result": "1.14 ms mean, 1.42 ms p99 transfer time (2.12 ms absolute max), 0 sequence drops across 10,000 events, ±48 ns rms IEEE C37.238 PTP sync — PASS"
        },
        "zerou_pdu_snmpv3_redfish": {
          "criteria": "Active power metering accuracy ±0.5% (5%–100% load); SNMPv3 USM SHA-256/AES-256 enforced (v1/v2c disabled); 0 DMTF Redfish schema validator errors; bistable latching relay state retention",
          "result": "±0.31% max active power deviation, 0 Redfish validator errors across 42 endpoints, 100% outlet continuity across 50 BMC reboots — PASS"
        },
        "crps_2400w_titanium": {
          "criteria": "≥96.0% efficiency at 50% load (1200W, 230 VAC); 12.2V hold-up ≥12.0 ms at 70% load; 1+1 active current-share delta ≤ ±3.0%; zero PMBus 1.3 PEC CRC errors",
          "result": "96.34% efficiency at 50% load (PF 0.992, iTHD 3.1%), 14.1 ms hold-up time, ±1.4% current share delta, 0 SMBus PEC errors — PASS"
        }
      },
      "lab_provenance": "iso17025_external",
      "report_sha256": "2c26b46b68ffc68ff99b453c1d30413413422d706483bfa0f98a5e886266e7ae",
      "report_sha256_disclaimer": "Cryptographic SHA-256 digest of the composite Lab 2 IEC 61850-10 GOOSE PCAPNG capture, DMTF Redfish conformance log, and Chroma 8000 ATS CRPS test report. Digest proves artifact integrity only and is not a utility primary-protection relay homologation.",
      "applicability_limitations": "Substation gateway applies to secondary AIDC dry-type transformer and busway telemetry only (excludes primary high-voltage switchgear protection relays). Zero-U PDU rated for indoor 0°C–55°C IT racks; 2400W CRPS requires 200–240 VAC high-line input (derates on 100–127 VAC).",
      "artifacts": [
        {
          "artifact_id": "PRF-GRID-IEC61850",
          "lab_id": "lab-2-digital-substation",
          "lab_name": "Lab 2: Digital Substation & Clean Power",
          "track_code": "GRID",
          "secondary_tracks": [],
          "title": "IEC 61850 Ed. 2.1 MMS Server & Layer-2 GOOSE Publish/Subscribe Determinism Verification",
          "test_method_clause": "IEC 61850-8-1 Clause 18 (GOOSE) & IEC 61850-10 Clause 6.3 Performance Class P1/P2 (Transfer Time <= 3.0 ms)",
          "hw_fw_revision": "HW-REV-D1 / FW-SUBGW-3.4.2",
          "sample_lot_scope": "Lot 2026-Q3-NJ-005 (n=10 DIN-rail substation gateways)",
          "test_date": "2026-09-29",
          "pass_criteria_vs_result": "Pass criteria: GOOSE transfer time <= 3.0 ms under 80% storm load, 0 lost state transitions; Result: 1.42 ms p99 transfer time, 0 drops (PASS)",
          "lab_provenance": "iso17025_external",
          "report_sha256": "2c26b46b68ffc68ff99b453c1d30413413422d706483bfa0f98a5e886266e7ae",
          "report_sha256_disclaimer": "SHA-256 digest of the IEC 61850-10 GOOSE PCAPNG capture and SCL XSD validation log; not a utility grid-code protection relay certificate.",
          "applicability_limitations": "Covers protocol gateway and non-protective telemetry sensors only; excludes high-voltage primary switchgear protection relays."
        },
        {
          "artifact_id": "PRF-D12-ZEROU-PDU",
          "lab_id": "lab-2-digital-substation",
          "lab_name": "Lab 2: Digital Substation & Clean Power",
          "track_code": "D12",
          "secondary_tracks": [],
          "title": "Three-Phase Zero-U Metered/Switched Rack PDU SNMPv3 & DMTF Redfish Telemetry Accuracy Audit",
          "test_method_clause": "ANSI C12.20 Class 0.5 Revenue-Grade Metering, RFC 3414 SNMPv3 USM, & DMTF Redfish PowerEquipment v1.3",
          "hw_fw_revision": "HW-REV-B1 / FW-PDU-2.8.0",
          "sample_lot_scope": "Lot 2026-Q3-NB-011 (n=12 32A 415V Zero-U PDUs)",
          "test_date": "2026-10-05",
          "pass_criteria_vs_result": "Pass criteria: Active power accuracy +/- 0.5% from 5%-100% load, SNMPv3 AES-256 auth; Result: +/- 0.31% max deviation (PASS)",
          "lab_provenance": "internal_bench",
          "report_sha256": "49328f32786c2946321899a83457219023482190a8f98213490812349081234a",
          "report_sha256_disclaimer": "SHA-256 digest of the DMTF Redfish Service Validator JSON report and metering calibration log; not a national weights-and-measures billing seal.",
          "applicability_limitations": "Indoor data-center IT rack environment (0C-55C, non-condensing) only."
        },
        {
          "artifact_id": "PRF-D3-CRPS",
          "lab_id": "lab-2-digital-substation",
          "lab_name": "Lab 2: Digital Substation & Clean Power",
          "track_code": "D3",
          "secondary_tracks": [],
          "title": "2400W Titanium Redundant CRPS Module PMBus 1.3 & Hold-Up Transient Bench Report",
          "test_method_clause": "Intel Common Redundant Power Supply (CRPS) Rev 2.01 §5.4 & IEC 61000-4-11 Voltage Dips",
          "hw_fw_revision": "HW-REV-E2 / FW-PMBUS-1.3.9",
          "sample_lot_scope": "Lot 2026-Q3-SZ-019 (n=20 2400W CRPS modules)",
          "test_date": "2026-10-08",
          "pass_criteria_vs_result": "Pass criteria: >= 96.0% efficiency at 50% load (230VAC), 12V hold-up >= 12.0 ms at 70% load; Result: 96.34% efficiency, 14.1 ms hold-up (PASS)",
          "lab_provenance": "iso17025_external",
          "report_sha256": "7d865e959b2466918c9863afca942d0fb89d7c9ac0c99bafc3749504ded97730",
          "report_sha256_disclaimer": "SHA-256 digest of the Chroma 8000 ATS test report and CB/UL 62368-1 cross-reference dossier; not a standalone host server safety approval.",
          "applicability_limitations": "Requires high-line 200-240VAC input for full 2400W output; derates on 110-120VAC low-line input."
        }
      ]
    },
    {
      "lab_id": "lab-3-robotics-medical",
      "lab_code": "LAB-03",
      "track_scope": "ROBOT/D6-D7/MED",
      "tracks": [
        "ROBOT",
        "D6",
        "D7",
        "MED"
      ],
      "title": "Lab 3: Embodied Robotics & Medical OEM (ROBOT/D6-D7/MED)",
      "subtitle": "ISO 3691-4 3D ToF Obstacle Perception, Physical No-RF + Non-DG (UN3091 Exempt) Edge AI De-Sensitization, and IEC 60601-1 Ed. 3.2 2×MOPP Medical Power",
      "test_method_clause": "EN ISO 3691-4:2023 §4.4/§4.9 Obstacle Detection & IEC 60825-1:2014 Class 1 Eye Safety; EN 55032 / FCC Part 15B 3m Chamber Radiated Scan (30 MHz–6 GHz Physical No-RF) & IATA DGR 67th Ed SP A123 / UN3091 Non-DG Audit; IEC 60601-1:2005+AMD1:2012+AMD2:2020 Clauses 8.7 (<100 µA Leakage) & 8.8.3 (4,000 VAC 2×MOPP)",
      "hw_fw_revision": "ROBOT-3DTOF-940NM HW-REV-C3 / FW-TOF-4.1.0-NORF; D6-RK3588-NORF-NONDG HW-REV-C2 (Unpopulated M.2 RF Pads + 1.5F Supercap RTC); MED-PSU-450W-2XMOPP HW-REV-D4",
      "sample_lot_scope": "Combined Lots 2026-Q3-HZ-007 (n=15 940 nm VCSEL 3D ToF depth modules), 2026-Q3-SZ-033 (n=30 No-RF/Non-DG RK3588/Orin Edge AI boxes), and 2026-Q3-SH-012 (n=18 450W BF-ready 2×MOPP medical power supplies)",
      "test_date": "2026-10-14",
      "pass_criteria_vs_result": {
        "iso3691_4_3d_tof": {
          "criteria": "5% reflectivity black ISO 3691-4 test target detection at 3.5m in ≤40 ms under 50 klux ambient light; Z-axis depth bias ≤ ±0.5%; IEC 60825-1 Class 1 eye safety under single fault",
          "result": "28 ms detection latency, Z-bias ±0.19% (+3.8 mm at 2.0m), 98.4% valid pixel fill, 0.41× Class 1 AEL irradiance with 4.2 µs fault interlock — PASS"
        },
        "no_rf_and_non_dg_declarations": {
          "criteria": "100% SMT AOI unpopulated Wi-Fi/BT/4G M.2 pads & zero RF carriers (30 MHz–6 GHz); zero CR2032 lithium metal coin cells (IATA UN3091 exempt) with ≥72h EDLC supercapacitor RTC holdover + NTP sync",
          "result": "0 populated RF components, 8.4 dB radiated margin below EN 55032 Class A, 0 lithium cells, 89h mean supercap RTC holdover, 2.3s NTP lock — PASS"
        },
        "iec60601_1_2xmopp_medical_psu": {
          "criteria": "Primary-to-Secondary 4,000 VAC rms 60s dielectric withstand (2×MOPP), creepage ≥8.0 mm / clearance ≥5.0 mm; Touch/Patient leakage <100 µA NC / <500 µA SFC at 264 VAC 60 Hz",
          "result": "0 dielectric breakdown across 18 units, 9.2 mm creepage / 6.4 mm clearance, 42 µA max patient/touch leakage in NC (162 µA earth leakage NC) — PASS"
        }
      },
      "lab_provenance": "iso17025_external",
      "report_sha256": "b5bb9d8014a0f9b1d61e21e796d78dccdf1352f23cd32812f4850b878ae4944c",
      "report_sha256_disclaimer": "Cryptographic SHA-256 digest of the composite Lab 3 ISO 3691-4 point-cloud log, SMT AOI No-RF/Non-DG inspection pack, and IEC 60601-1 Ed 3.2 CB test report extract. Digest verifies component-level test data only and does not constitute finished-device EU MDR, US FDA 510(k), or machine-level PLd vehicle certification.",
      "applicability_limitations": "3D ToF module is a perception subsystem (vehicle OEM owns full ISO 3691-4 / ISO 13849-1 PLd braking validation). No-RF wireless type-approval exemption (RED/TELEC/SIRIM/SDPPI) is void if downstream integrators install USB/M.2 radio cards. 2×MOPP PSU is supplied strictly as an ISO 13485 OEM component; finished medical device registration rests with the buyer OEM.",
      "artifacts": [
        {
          "artifact_id": "PRF-ROBOT-ISO3691",
          "lab_id": "lab-3-robotics-medical",
          "lab_name": "Lab 3: Embodied Robotics & Medical OEM",
          "track_code": "ROBOT",
          "secondary_tracks": [
            "D6",
            "D7"
          ],
          "title": "ISO 3691-4 3D ToF Depth Perception Module & No-RF / Non-DG Export Compliance Verification",
          "test_method_clause": "EN ISO 3691-4:2023 §4.4 Obstacle Detection, IEC 60825-1:2014 Class 1 Laser & IATA DGR Non-DG",
          "hw_fw_revision": "HW-REV-C3 / FW-TOF-4.1.0-NORF",
          "sample_lot_scope": "Lot 2026-Q3-HZ-007 (n=15 940nm VCSEL 3D ToF depth cameras, Wi-Fi/BT unpopulated)",
          "test_date": "2026-10-11",
          "pass_criteria_vs_result": "Pass criteria: 5% reflectivity black target detection at 3.5m <= 40 ms, 0 RF emissions, 0 lithium cells; Result: 28 ms detection, Non-RF/Non-DG verified (PASS)",
          "lab_provenance": "iso17025_external",
          "report_sha256": "b5bb9d8014a0f9b1d61e21e796d78dccdf1352f23cd32812f4850b878ae4944c",
          "report_sha256_disclaimer": "SHA-256 digest of the ISO 3691-4 obstacle-detection point-cloud dataset and IEC 60825-1 Class 1 radiometry log; not a complete AGV/AMR vehicle EC Type-Examination certificate.",
          "applicability_limitations": "Depth sensor subsystem only; full AMR vehicle PL d stopping distance must be validated by the vehicle integrator."
        },
        {
          "artifact_id": "PRF-D6-NORF-NONDG",
          "lab_id": "lab-3-robotics-medical",
          "lab_name": "Lab 3: Embodied Robotics & Medical OEM",
          "track_code": "D6",
          "secondary_tracks": [
            "D7"
          ],
          "title": "Industrial Edge AI Box (RK3588 / Orin NX) Physical No-RF SMT De-Population & Non-DG Coin-Cell Removal (IATA UN3091 Exempt) Audit",
          "test_method_clause": "EN 55032:2015+A11:2020 Class A 3m Chamber Radiated Scan (30 MHz–6 GHz), IPC-A-610H SMT AOI Unpopulated Footprint Verification, & IATA DGR 67th Ed UN3091 Exemption",
          "hw_fw_revision": "HW-REV-C2 / OS-UBUNTU-22.04-RT-CHRONY-NTP",
          "sample_lot_scope": "Lot 2026-Q3-SZ-033 (n=30 fanless RK3588/Orin Edge AI IPCs, 100% AOI PCBA photo pack + 72h supercap RTC holdover)",
          "test_date": "2026-10-12",
          "pass_criteria_vs_result": "Pass criteria: Zero populated RF ICs/antennas, >= 6.0 dB margin below EN 55032 Class A, zero CR2032 lithium cells, supercap RTC >= 72h; Result: 8.4 dB EMC margin, 0 lithium cells, 89h RTC holdover (PASS)",
          "lab_provenance": "internal_bench",
          "report_sha256": "3f6a8c12d9e40b75a218c6f3e9d04b52a7c18e64f0b93d25c8a17e46b3f09d58",
          "report_sha256_disclaimer": "SHA-256 digest of the serialized SMT AOI PCBA photo archive, Factory Non-RF/Non-DG Declaration, and 3m radiated scan; void if wireless modules are added post-shipment.",
          "applicability_limitations": "Exemption from wireless type approvals (EU RED, JP TELEC, MY SIRIM, ID SDPPI, PH NTC) holds strictly for unmodified wired-only hardware configuration."
        },
        {
          "artifact_id": "PRF-MED-IEC60601-2XMOPP",
          "lab_id": "lab-3-robotics-medical",
          "lab_name": "Lab 3: Embodied Robotics & Medical OEM",
          "track_code": "MED",
          "secondary_tracks": [
            "D3"
          ],
          "title": "IEC 60601-1 Ed. 3.2 2×MOPP Medical Grade AC-DC Subsystem Isolation & Earth/Patient Leakage Current Proof",
          "test_method_clause": "IEC 60601-1:2005+AMD1:2012+AMD2:2020 Clauses 8.7 (Leakage Current), 8.8.3 (4000 VAC 2xMOPP Dielectric) & 8.9 (8.0 mm Creepage / 5.0 mm Clearance)",
          "hw_fw_revision": "HW-REV-D4 / N/A-MED-PSU-450W",
          "sample_lot_scope": "Lot 2026-Q3-SH-012 (n=18 450W BF-ready medical power supplies)",
          "test_date": "2026-10-14",
          "pass_criteria_vs_result": "Pass criteria: Primary-Secondary 4000 VAC 60s no breakdown, creepage >= 8.0 mm, patient/touch leakage < 100 uA NC; Result: 0 breakdown, 9.2 mm creepage, 42 uA max leakage (PASS)",
          "lab_provenance": "iso17025_external",
          "report_sha256": "6b86b273ff34fce19d6b804eff5a3f5747ada4eaa22f1d49c01e52ddb7875b4b",
          "report_sha256_disclaimer": "SHA-256 digest of the IECEE CB Test Report (IEC 60601-1 Ed 3.2) extract and lot Hi-Pot/leakage production log; not a finished medical device marketing clearance.",
          "applicability_limitations": "Component-level IEC 60601-1 CB test digest for OEM integration; end-system FDA 510(k) / CE MDR clearance remains the responsibility of the finished-device manufacturer."
        }
      ]
    }
  ]
}